Minimum pulse width decides whether your DCIR data is usable
A pulse shorter than the instrument response is not a pulse, it is an artefact. Here is how pulse width, acquisition rate and response time interact when you measure internal resistance.
NEWARE Application Engineering
The measurement problem
Internal resistance is inferred from a voltage step under a known current. The inference assumes the voltage you recorded is the voltage the cell produced. When the pulse is shorter than the time the instrument needs to settle, part of the recorded step is the tester's own lag, and the resistance value carries that error into every downstream calculation.
The practical consequence is that two labs can test the same cell, report different DCIR values, and both be internally consistent. The disagreement sits in the hardware specification, not the analysis.
Two laboratories can test the same cell, report different DCIR values, and both be internally consistent. The disagreement sits in the hardware specification, not the analysis.
Three numbers that must agree
Pulse width, current response time and data record frequency form a chain. If the pulse is 10 ms, the response time must be well inside that window, and the acquisition rate must place enough samples inside the pulse to resolve the step rather than average it away.
- Response time defines how fast the channel reaches the commanded current
- Record frequency defines how many samples describe the step
- Minimum pulse width defines the shortest event the system will execute faithfully
Where the series differ
Within the same manufacturer, pulse capability varies by more than an order of magnitude. A routine cell platform typically works in the 100 to 500 ms range, which is appropriate for capacity and cycle-life work. High-speed platforms reach 400 microseconds with 1000 Hz acquisition. High-current platforms sit in between, at roughly 10 ms with millisecond current response.
Choosing between them is not a matter of buying the fastest instrument. It is a matter of knowing the shortest event in your test plan and leaving margin below it.
A short checklist before you specify
Write down the shortest pulse in the plan, the number of samples you need inside it, and the ambient temperature range the test will run in. Then compare those three values against the datasheet rather than against a brochure claim. If the margin is thin, ask for a demonstration run on your own cell.